66
Instructions for Use, V2.0, RevG - 07/2016
Completing the program
• Once the parameters
EXACT
and
DIP
have been selected, the defini-
tion of your step is complete.
• Press
INSERT ROW
to create a new row for the next program step and
once again define all step parameters as described above.
• Keep creating program steps until your program is complete.
• Program steps whose definition is not correct (duplicate station as
-
signment, incorrect reagent,...) are highlighted in red in the
REAGENT
column. Such steps must be corrected accordingly.
• A program will not be ready to start until all details have been defined.
The last program step
Three different types of "last steps" can be defined:
1.
Last step = unload station.
Go to
CHOOSE STATION
and select
EXIT
as last station.
2.
Last step = transfer station to Cover Slipper (optional).
Go to
CHOOSE STATION
and select
CV
as last station.
The
REAGENT
column from the reagent list now allows selecting be-
tween CV program 1 to CV program 4.
The transferred rack is automatically processed with the selected pro-
gram (1 ... 4) in the CV5030.
3.
Last step = any station other than
EXIT
or
CV
.
Programs do not necessarily have to finish in the unload station, as a
particular staining protocol might e.g. have to end with a specific rea-
gent not contained in the last station.
Be sure to define the DURATION of the last step. Otherwise it will
not be carried out (default duration 00:00 will be assigned). Other-
wise, the program cannot be started.
5. Operation