37
Leica ST5020
5.3.3 Access levels
Administrator symbol
→
The Leica ST5020 may be configured to allow
two level of user access.
User access level
• Users may run programs and view results.
On this level, the
USER
symbol is displayed
in the upper right corner of the touch screen.
Supervisor access level
• Administrators may perform all user func
-
tions, and additionally create programs and
perform the instrument setup functions.
• To access supervisor level, press
SUPER-
VISOR
: the keyboard will be displayed. If a
password has been allocated, the keyboard
will appear on the screen.
(See
chapter 5.4.3
for details on password
allocation).
Enter the 'Admin' password and confirm.
The
SUPERVISOR
symbol will then be dis-
played instead of the
USER
symbol and the
SUPERVISOR
button will be replaced by the
USER
button.
User symbol
Fig. 31
In SUPERVISOR mode, NO programs
can be run. Switch to USER mode to
start a program.
You can only go back to SUPERVI-
SOR mode (e.g. to select / change
settings) once all running programs
have been finished.
5. Operation
The Leica ST5020 may be configured
so that all functions are accessible to
all staff - effectively the instrument is
in Supervisor access level the whole
time. See
chapter 5.4 "System setup"
.