64
Instructions for Use, V2.0, RevG - 07/2016
Assigning reagents
• If no reagent is assigned to a program step,
press
REAGENT
in the header row to switch
to the reagent list (see
chapter 5.4.7
).
Highlight one of the reagents in the list and
press
CONFIRM
to assign it to the program
step.
• The reagent is now assigned to the step but
not yet to the station in which the step must
be executed.
For this reason, the row is highlighted in red
(
1
) indicating that the program step is not yet
ready to be started.
Assign the reagent selected for the step to
the correct station in
INSTRUMENT STATUS
(see
chapter 5.5.2
).
This is required to ensure that the correct
reagent is in the station since the instrument
itself cannot detect it.
• Define the
DURATION
of the step, wheth-
er the step is
EXACT
or
NON-EXACT
and
whether or not there are
DIPS
.
Fig. 62
↓
Defining the duration of a step
• Press
DURATION
and enter the desired duration of the step. Press
CONFIRM
: the selected duration is displayed in the
DURATION
col-
umn.
Caution!
If no step duration is selected, the default duration "00:00:00" will be
used, i.e. the corresponding step will NOT be carried out.
1
↓
Fig. 61
5. Operation
↓