27
Leica ST5020
To do so, proceed as follows:
• Remove all vessels located in stations 1 - 24
so that the vessel holder frames can be used
to support the box level for measuring pur-
poses.
• All four instrument feet (
37
) are height-ad-
justable.
To adjust, place single head wrench, size 13
(standard delivery) onto hexagon bolt (
36
)
and adjust each foot as needed.
Make sure, all four instrument feet
are firmly set on the bench surface
once the instrument adjustment has
been completed.
36
37
Adjustable
instrument feet
Fig. 21
Bring the instrument
to an exact horizontal
position by screwing the
adjustable feet in / out as
needed.
•
For leveling, position box lev-
el (
38
)
alternately
in positions
(
A
and
B
) - see
Fig. 22
on the
vessel holder frame.
• Screw the instrument feet
inward / outward as needed,
gradually bringing the instru-
ment into a horizontal posi-
tion. Start with the instrument
foot located opposite the
highest point of the labora-
tory bench.
• Finally, place the box level in
positions (
C
und
D
) to double-
check whether the instru-
ment is in fact in a horizontal
position.
Fig. 22
Fig. 22
shows all four points (A to D) where the box level
must be positioned for leveling the instrument.
For leveling, the box level must be placed onto vessel
holder frame (39). - Do not place the box level elsewhere.
38
39
39
A
B
C
D
4. Setting up the instrument