5.5.4 Change station type
Load/unload station
→
Reagent station
• Load stations
L33
-
L35
and unload stations
E26
-
E28
can be defined as
reagent stations, if needed.
• Both station types must be redefined station by station (one station at
a time), starting from the 'center'; i.e., if you wish to convert load and/
or unload stations into reagent stations, you must start by converting
stations
E28
or
L33
, respectively.
• The
CHANGE STATION TYPE
button (
5
) only becomes available to the
stations located further towards the outside after stations
E28
, or
L33
,
respectively, have been converted into reagent stations.
The same, in reverse order, applies for converting stations back into
load / unload stations.
• Press
CHANGE STATION TYPE
(
5
)
to access the
STATION TYPE
CHOICE
menu.
Proceed further as described in
chapter 5.5.3
.
Fig. 54
5
Tap water station
→
Reagent station
• It is also possible to convert the water stations
W7
-
W12
into reagent
stations, if needed. No particular order must be followed in that case.
• Follow the same procedure as described above. In addition, however,
the yellow tap water containers must exchanged for reagent contain-
ers to prevent reagents from spilling.
If the RMS is on, all reagent stations will be registered, including
those that have been converted into reagent stations later.
↓