75
Leica ST5020
• When pressing
YES
a second prompt opens up, where you need to
confirm that you have removed the rack from the station it was located
in last.
This step is necessary as the instrument itself is not able to detect
whether there is a rack in that station or not.
Wait until the transfer arm has finished the movement in progress
and is located in the parking position (rear left). Open the lid and
remove the rack. WAIT until the station number is displayed on the
screen - once this message is being displayed, confirm that the
rack has been removed.
• Press
CONFIRM
- the current staining run is aborted and removed from
the list of pending runs. Staining will be resumed starting with the next
rack.
If the program highlighted by the blue bar has already been comple-
ted (rack already located in unload station or just being transferred to
unload station), the CANCEL RACK button is disabled.
Cancelling racks
5. Operation
↓
↓
Racks can be cancelled if needed.
• Press
STAINING SCHEDULE
to access the stain-
ing schedule menu.
• Select the rack you wish to cancel (highlight
with blue bar).
• Press
CANCEL RACK
.
• Press
YES
to confirm.
If
NO
is pressed, the instrument will switch
back to the staining schedule menu without any
changes being made.
↓
Fig. 71