74
Instructions for Use, V2.0, RevG - 07/2016
Removing racks from stations other than unload
If a program finishes in a station other than unload, alarm type 3 sounds
at the end of the program and there will be a screen prompt requiring the
user to remove the rack.
• Press
CONTINUE
to confirm that the rack has been removed from the
station.
This is necessary, as the instrument will otherwise not be able to rec-
ognize whether there is still a rack in that station or not.
• Confirmation of rack removal also switches off the alarm. The stain
-
ing program is considered to be finished and removed from the list of
pending staining runs.
5. Operation
5.8.4 Finishing programs
Unloading racks from the unload drawer
If a rack is located in the unload drawer, the alarm type 1 sounds.
• The code of the completed program and (in the bottom row) the basket
number are displayed in the unload station symbol. The colored pro-
gram indication blinks in two-second intervals.
• Carefully open the unload drawer and remove the rack. You may also
remove the entire reagent container together with the rack and re-
place it by another container.
↓
• Close the unload drawer - program code and rack
number must disappear from the station button.
• If, despite the unload drawer having been opened
and closed several times, the instrument still dis-
plays code and rack number of the rack already re-
moved, the transponder or the automatic recognition
system might be defective.
Press
CANCEL RACK
to manually 'unload' the clip.
Fig. 70