6. Cleaning and maintenance
6.1 Cleaning the instrument
Prior to cleaning the instrument, always switch off mains and un-
plug the power cord!
When handling cleaning detergents, follow the instructions of the
manufacturer and make sure all laboratory regulations in force in
your country are complied with.
Dispose of used reagents according to the laboratory regulations
in force in your country!
Immediately wipe away spilled solvents (reagents)! Solvent re-
sistancy of the lid surfaces is limited if exposed to solvents for a
lengthy period of time!
Do not use any of the following for cleaning the outside surfaces
of the instrument: alcohol, detergents containing alcohol (window
cleaner!), abrasive cleaning powders, solvents containing acetone
or xylene!
No liquid may come into contact with the electrical connections
or spill into the interior of the instrument and/or the transport arm!
Interior
Remove the reagent and wash vessels. Clean the stainless-steel inner
walls with a regular cleaner and rinse with water.
Transfer arm
The transfer arm contains sensitive electronic components. Therefore do
not clean with any liquid but simply wipe down with a slightly moistened
cloth.
Outside surfaces
Clean the painted outside surfaces with a mild detergent and subsequent-
ly wipe down with a moistened cloth.
Do not use any solvents for cleaning the outside surfaces and/or the lid!
TFT touch screen
Use a commercial screen cleaner for cleaning the touch screen. Use as
directed by the manufacturer.