Operating Manual Leica EM ACE600
Version 01/19
Page 31
The cover in the middle of the stage could be in the way.
Loosen the screw with an Allen key and lift up the cover slightly, the stage will
slide in.
Do not forget to change the quartz settings from stage to side (see 3.1.1).
Minimum required working distance is 80mm and the measurement is calibrated
to 0 tilt.
Note for sputter processes. When using the side settings for the quartz
Argon pressure of 2x10-2 to 9x10-3 mbar or lower should be used for the
sputtering. Above the measurement is less accurate.