Operating Manual Leica EM ACE600
Version 01/19
Page 91
Wipe around the sealing surface with a lint free tissue and ethanol
Reattach the source cover, tighten the evaporation head screws and close the
cover.
4.9.2 Choosing a carbon rod protocol
When the head is inserted after completing the preparation in 4.9.1 an
evaporation process can be run. Select evaporation on the first screen to enter
the process screen. In the library of recipes (1) all stored protocols are available
and can be chosen. Tapping on the characteristics of the protocol (2) opens the
list of recipes. There the parameters for the process can be changed and saved.