Operating Manual Leica EM ACE600
Version 01/19
Page 121
E2003
Process
Process terminated mains
supply voltage out of range
150<>210 V
Process - aborted. Mains supply
voltage out of range! (150<>210
V) Check if right voltage is
delivered by the network
E2004
Process
Process terminated, Mains
supply voltage out of range
>250 V
Process stop. Mains supply
voltage out of range! (>250 V)
Check if right voltage is delivered
by the network
E2005
Process
Process terminated, Current
not reached, Check target
and sputter source
Process aborted - Unable to reach
sputter-current. Check if pressure
is valid or wrong or no target is
used.
E2007
Process
Process terminated, Line
voltage fluctuations
Process aborted - Line voltage
fluctuations too big!
E2009
Process
Process terminated, Sputter
source short circuit! Check
target and sputter source.
Process aborted - Short-circuit
because of invalid assembly of
head, ring or target. Check
deposited solid material on shutter
E2010
Process
Process terminated, ignition
fault, Check argon flow,
Check target and sputter
source
Process aborted - Unable to
establish plasma ignition.
E2011
Process
Process terminated, Sputter
current could not be
stabilized
Process aborted - Unstable
plasma current. Unclean target
could be the reason, clean with
isopropanol and try again.
E2012
Process
Process terminated, High
voltage board defect, please
contact service
Process aborted - High voltage
board defect > Change board,
since output cannot be controlled
anymore
E2018
Process
Operation Response Timeout
The control unit cannot stabilize
the voltage in the sputter head.
W2019
Warning
Purge Base-Vacuum (1.0E-4
Mbar) Timeout. Process
Anyway?
Begin Sputter using Standby
Mode
Glow discharge specific:
E2109
Process
Process terminated, Short
circuit in glow-discharge,
Check shutter panels
Process aborted - Short-circuit
because of invalid assembly of
plates. Check deposited solid
material on shutter