Operating Manual Leica EM ACE600
Version 01/19
Page 53
4.6.3 Carbon thread materials
New carbon thread materials can be defined to adjust the pulsing or flashing
characteristics. Open Menu, tab file and material management. Select an existing
material or add a new one. Set the parameters for the degassing process (heat
current, heat voltage and heat time). Pulse power is the value the instrument tries
to reach (adjusting current and voltage accordingly). Time between the pulses and
duration of a pulse is defined.
Flash current and voltage are a limit. The flash voltage can be adjusted to desired
value.
1