Operating Manual Leica EM ACE600
Version 01/19
Page 118
W0010
Warning
Turbo temperature too high,
Service diaphragm pump
Turbo temperature over-heat. Stop
pumping and wait.
W0011
Warning
Critical disk space, Back-up
files before clean up?
Not enough memory left on
device. Clean up device > Delete
old logs, screenshots and update
folders.
E0012
Error
No pump calibration table
found, please calibrate pump
to ensure correct behaviour
No pump calibration file found.
Calibrate pump in service.
E0013
Error
No available coating method
detected, Restart system
No available coating method
controller detected. Check
communication cables and HiVac
controller status.
W0014
Warning
Service interval gauge
exceeded, Service gauge
Gauge service interval exceeded.
Reset interval in software after
service.
W0016
Warning
Gauge pressure out of
bounce, Restart system
Invalid gauge pressure value
received. Check gauge connection
or restart controller.
E0017
Error
Sudden Vacuum Break
Down; Check Valves and
Device For Leaks
Vent after the Valve, then Open
the FSA-Valve, Pressure Drop
E0018
Error
Vacuum Timeout; Check
Valves and Device For Leaks
W0019
Warning
Side Stage, Working
Distance Lower than 80 mm
HiVac Controller Specific:
W1300
Warning
No quartz detected
Quartz not detected. Warning only
active in debug mode.