Page 120
W1904
Process
Process terminated, head (1)
over-temperature
Process aborted - Head-
temperature overheated! Temp. >
65°C
W1905
Process
Process terminated, head (2)
over-temperature
Process aborted - Head-
temperature overheated! Temp. >
65°C
E1100
Error
No gauge detected, Restart
system
No gauge detected at start-up of
software. Restart device.
Carbon Thread Controller Specific:
W4001
Warning
No thread available
No threads detected. Abort
process only after resistance
calculation.
W4002
Warning
All threads exhausted
Process stopped - All threads
exhausted in process, but still
some pulses to proceed (or
thickness not reached) > Abort
process
W4003
Warning
# of threads available, # of
flashes not possible
Process stopped - Amount of
existing threads smaller than
desired number of flashes
E4011
Process
Process terminated, no
power detected, Check door
and cover
Process aborted - Carbon Thread
supply not powered. Check door
and cover sensors
E4012
Process
Process terminated, Main
supply voltage fault, Check
door and cover
Process aborted - Carbon Thread
mains relay not switched on.
Check door and cover sensors
E4013
Process
Thread Measuring Timeout;
Restart Process
E4014
Process
Supply Off Response
Timeout
Sputtering Controller Specific:
E2001
Process
Process terminated mains
supply voltage fault, check
door and cover
HV-board not supplied. Check
door and cover sensors
E2002
Process
Process terminated, mains
supply voltage out of range
<85 V
Process - aborted. Mains supply
voltage out of range (<85 V)!
Check if right voltage is delivered
by the network