Operating Manual Leica EM ACE600
Version 01/19
Page 9
2.4
Safety information
2.4.1 General instructions
Covers protect all electronic components: door lock (1), source cover (2), and
housing (3). The door and source covers are equipped with sensors that cut off
power supply when opened. Only authorized Leica representatives are allowed to
remove the housing. Additionally, there are software switches, which cut off power
when a malfunction is noticed.
The coating sources are protected against overheating. If at any time, the source
temperature reaches 65° C, the system automatically pauses the current process
only to continue after the temperature has sank below 45° C.
In case of a sudden and unexpected vacuum break, the instrument switches off
automatically to protect the pump and electric parts.
All electrical parts are safely separated from the user by the front cover and back
housing.
If the Leica EM ACE600 coating system is damaged or malfunctions, all use of the
system should be suspended until the issue has been corrected.
All modifications and conversions to the system are prohibited and invalidate the
warranty.
3
2
1