![LeCroy WaveRunner 6Zi Getting Started Manual Download Page 94](http://html1.mh-extra.com/html/lecroy/waverunner-6zi/waverunner-6zi_getting-started-manual_1866372094.webp)
Getting Started Manual
WR6Zi-GSM-E
84
A trigger is generated when the condition is met after this number,
counted from the last trigger. The count is restarted on each trigger. For
example, if the event number is two, the trigger will occur on the third
event.
Edge Trigger with Holdoff by Events (in this example, two events). The bold
edges on the trigger source indicate that a positive slope has been selected.
The broken, upward-pointing arrows indicate potential triggers, while the bold
ones show where triggers actually occur after the holdoff expires.
Auxiliary Input Trigger
Some instrument models provide auxiliary input trigger capability. It's
done as a pattern trigger, on the
Ext
dialog.
Select
X1
, or
÷10
. You can also select from
DC50Ω
,
Gnd
, and
DC1MΩ
Impedance