![LeCroy WaveRunner 6Zi Getting Started Manual Download Page 70](http://html1.mh-extra.com/html/lecroy/waverunner-6zi/waverunner-6zi_getting-started-manual_1866372070.webp)
Getting Started Manual
WR6Zi-GSM-E
60
2.
Click the
Sequence
tab.
3.
Under
Acquisition Settings
, touch inside the
Num Segments
data
entry control and provide a value using your preferred input
control method - for more information regarding the touch screen
controls, refer to
Dialog Area
(on page 46).
Note
: The number of segments you choose to display can be less
than the total number of segments in the waveform. For
example, in the pop-up images above, the number of display
segments is 10, but the total number of segments entered in the
timebase dialog's
Num Segments
control is 100.
4.
Touch the
Enable Timeout
checkbox.
5.
Touch inside the
Timeout
data entry control and provide a
timeout value.
Note
: Use the sequence mode timeout to automatically interrupt the
sequence acquisition if the timeout value is exceeded without a valid
trigger. The timeout period accounts for instances when a
Num
Segments
miscount occurs for some reason and the oscilloscope waits
indefinitely for an unforthcoming segment. During that time, no
oscilloscope functions are accessible. By means of a timeout value,
however, the acquisition will be completed, the waveform displayed,
and control of the oscilloscope returned to the user after the timeout
has elapsed.
5.
Under
Display Settings
, touch inside the
Display mode
control,
and select a sequence mode display from the pop-up menu.
6.
Touch the S
INGLE
trigger front panel button.
Note
: Once a single acquisition has started, you can interrupt
the acquisition at any time by pressing the
S
INGLE
front panel
button a second time or by pressing the
S
TOP
front panel button.
In this case, the segments already acquired will be retained in
memory.