WaveRunner 6 Zi Oscilloscopes
33
WR6Zi-GSM-E
Note
: AUX INPUT and AUX OUTPUT connections are also typically
located on the Front Panel. Refer to the
Auxiliary Output Signals
(on
page 169) topic for
Auxiliary Output
dialog details.
Probe Interfaces
Probe Interfaces Overview
LeCroy oscilloscopes utilize one or more proprietary probe interfaces
providing a complete measurement solution from probe tip to
oscilloscope display. Compared to standard BNC and Probe Ring
interfaces, this intelligent interconnection between your instrument and
the probe or accessory eliminates the guesswork and errors that occur
when probe settings are made manually, and also offers the following
important signal fidelity advantages:
Upon connection to the oscilloscope channel, the probe is
recognized and some setup information, such as input coupling
and attenuation, is performed automatically.
System (probe plus oscilloscope) gain settings are automatically
calculated and displayed based on the probe attenuation.
Active probes typically provide automatic matching of probe to
oscilloscope response using probe response data stored in an on-
board EEPROM. This ensures the best possible combined probe
plus oscilloscope channel frequency response without requiring
the user to perform any de-embedding procedure.
Probe interfaces differ in bandwidth, so the interfaces contained on your
oscilloscope depend on the bandwidth rating of the oscilloscope model
you purchased.