These indicators have a long and short form, respectively.
The long and short forms of trace descriptor indicators.
Besides channel traces, math and parameter measurement labels are
also displayed. Labels are displayed only for traces that are turned on.
Vertical and horizontal trace descriptor (labels) are displayed below the
grid. They provide a summary of your channel, timebase, and trigger
settings. Make vertical or horizontal channel adjustments by touching
the respective label. The setup dialog for the function is shown beneath.
TimeBase
trace descriptor labels show the
trigger delay setting
,
time per
division
, and
sampling information.
Trigger
trace descriptor labels show the trigger mode (Auto, Normal, or
Stopped). It also shows the coupling (DC), trigger type (Edge), source
(C1), level (0 mV), and slope (Positive).
Setup information for horizontal cursors, including the time between
cursors and the frequency, is shown beneath the TimeBase and Trigger
trace descriptor labels.