WaveRunner 6 Zi Oscilloscopes
119
WR6Zi-GSM-E
1.
On the menu bar, touch
Measure
, then one of the Measure
Mode buttons:
Std Vertical
,
Std Horizontal
, or
My Measure
.
2.
Touch the
Histicons
checkbox to display thumbnail histograms
below the selected parameters.
Note
: For measurements set up in My Measure, you can quickly display
an enlarged histogram of a thumbnail histogram by touching the Histicon
you want to enlarge. The enlarged histogram appears superimposed on
the trace it describes. This does not apply to
Std Vertical
or
Std
Horizontal
measurements.
Persistence Histogram
You can create a histogram of a persistence display also by cutting a
horizontal or vertical slice through the waveform. You also decide the
width of the slice and its horizontal or vertical placement on the
waveform.
Note
: This math operation is different than the Histogram math
operation and is not affected by
Center
and
Width
settings made on any
existing Histograms.
P
ERSISTENCE
H
ISTOGRAM
S
ETUP
1.
Touch
Math → Math Setup...
on the menu bar.
2.
Either directly touch an
Fx
button on the Math dialog and select
your Math Operator, or touch an
Fx
tab to show its
corresponding dialog and touch the Operator1 dialog to select
your Math Operator. Select
Phistogram
from the
Select Math
Operator
pop-up.
3.
With your math operator selected and the Fx dialog shown,
touch inside the
Source1
field and select a source from the pop-
up.