![LeCroy WaveRunner 6Zi Getting Started Manual Download Page 123](http://html1.mh-extra.com/html/lecroy/waverunner-6zi/waverunner-6zi_getting-started-manual_1866372123.webp)
WaveRunner 6 Zi Oscilloscopes
113
WR6Zi-GSM-E
Characteristic
Track
Trend
Time Correlation to
Other Data
Yes
No
Monitors an
Evolution in the
Frequency Domain
Yes
No. Trend points are not
evenly spaced in time and
therefore cannot be used
for an FFT.
Monitors the
Evolution of a
Measurement
Parameter over
Several
Acquisitions
No. Track resets
after every
acquisition.
Yes
Ensures No Lost
Measurement Data
Yes. Maximum
time period that
can be captured
is limited by
acquisition
memory and
sampling rate.
No. Since data can be
accumulated over many
acquisitions, and since the
oscilloscope takes time to
calculate measurement
values and to display data
before the trigger is re-
armed, data can be
missed.
In general, Track is the tool to use if you want to capture a continuous
stream of data spaced closely together. To understand the change in a
parameter with time, Trend can be used if your data is spaced widely
apart and longer than the dead-time of the oscilloscope between
acquisitions. Think of Trend as a strip chart recorder for your
oscilloscope.
Creating and Viewing a Histogram
Histograms are graphical representations of data which divides it into
intervals or bins. These intervals/bins are plotted on a bar chart such that