112
math traces. A common usage for track is to observe the
modulation of a signal, such as amplitude, frequency, or pulse
width modulation. See
Creating a Track View
(on page 121).
Trends
provide a view of the measurement parameter over an
extended period of time and over multiple acquisitions. See
Creating and Viewing a Trend
(on page 120).
Pass/Fail Testing
(on page 122) allows definition of the normal or
abnormal measurement parameter value with a definable
pass/fail action alerting the operator when a condition is met.
WaveScan
allows searching through a single acquisition for a
measurement parameter that meets a variety of conditions, or
scanning through multiple acquisitions with allowable trigger
actions when conditions are met. In addition, a variety of views
can be used to understand the specific behaviors of searched or
scanned measurement parameters. See
Introduction to
WaveScan
(on page 139).
These capabilities are only possible with the measurement of every
occurrence of a measurement parameter in a long acquisition.
Both Track and Trend are tools that can be used to plot measurement
data and observe variations with respect to time. Differences between
Track and Trend are summarized in the following table:
Characteristic
Track
Trend
Representation
Parameter value
vs. time
Parameter value vs. event
Behavior
Non-cumulative
(resets after
every
acquisition).
Unlimited
number of events
Cumulative over several
acquisitions up to 1
million events