WaveRunner 6 Zi Oscilloscopes
127
WR6Zi-GSM-E
Setups are conveniently shown in the Summary box of the
Qx
dialog as
follows (shown on the Dual Param Compare right-hand dialog):
1.
Touch
Analysis → Pass/Fail Setup...
from the menu bar.
2.
Touch one of the
Qx
tabs and a setup dialog for the position is
shown.
3.
Touch inside the
Condition
field in the main dialog and select
DualParamCompare
.
4.
Touch inside the
Source1
and
Source2
fields and select a source
from the pop-up menu.
5.
Touch inside the
ParamCompare
mini-dialog field and select a
source from the pop-up menu.
6.
Touch inside the
Compare Values
field and select
All
or
Any
from
the pop-up menu.
By selecting
All
, the test are true only if every waveform falls
within the set limit. Selecting
Any
makes the test true if just one
waveform falls within the limit.