No manual voltage selection is required because the instrument
automatically adapts to line voltage.
Power Consumption
Power Consumption (Nominal): 400 Watts (400 VA)
Max Power Consumption (with all PC peripherals, active probes connected to 4
channels, and MSO active): 500 Watts (500 VA)
Power consumption (in Standby Mode): 15 Watts
Power and Ground Connections
The instrument is provided with a 10A/250V 18AWG rated grounded
cord set containing a molded three-terminal polarized plug and a
standard IEC320 (Type C13) connector for making line voltage and safety
ground connections. The AC inlet ground terminal is connected directly
to the frame of the instrument. For adequate protection against
electrical shock hazard, the power cord plug must be inserted into a
mating AC outlet containing a safety ground contact. Ensure that the
disconnect device is readily accessible.
WARNING - Electrical Shock Hazard
Only use the power cord provided with your instrument.
Any interruption of the protective conductor inside or outside of the
oscilloscope, or disconnection of the safety ground terminal creates a
hazardous situation. Intentional interruption is prohibited.
In Standby mode, the oscilloscope is still connected to the AC supply. The
instrument can only be placed in a complete Power Off state by flipping
the main power switch on the rear of the instrument (if provided) to the
off (Zero) position or by removing the power cord. In cases where no
main power switch is provided, removing the main power cord is the
disconnect device.