WaveRunner 6 Zi Oscilloscopes
11
WR6Zi-GSM-E
Changes can be made to the oscilloscope's original factory settings via
the
Power Options Properties
menu in Windows under
Control Panel →
Power Options
. A Windows Power Option named
Standby
provided
control of only the oscilloscope’s computer subsystems (CPU, hard drive,
etc.) and did not affect the other subsystems within the oscilloscope. In
general, these other subsystems remain fully powered. For additional
information on setting these Power Options, see the Windows Help
menu or other related technical documentation. In terms of control
buttons, this oscilloscope uses only a power button/switch and therefore
references to a sleep button are not applicable.
The oscilloscope can always be placed in the Standby state (LED Off) –
Power Off (except for some housekeeping circuits) by pressing and
holding in the Standby toggle switch for approximately 5 seconds.
CAUTION
Hibernate mode is not supported by the Power button. Therefore,
Windows®
Power Options
must not be changed from the default
Never
setting to
System Standby
or
System Hibernate
modes.
CAUTION
Always use the
Power Switch
or
File → Shutdown
command in the
software to properly shut down your instrument. Never substitute this
practice by pulling the instrument power cord from the socket or
shutting down a connected power strip.
Calibration
The recommended calibration interval is one year. Calibration should be
performed by qualified personnel only.
Cleaning
Clean only the exterior of the instrument, using a damp, soft cloth. Do
not use chemicals or abrasive elements. Under no circumstances allow
moisture to penetrate the instrument.