Ch.3 Settings
< 65 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
・
・
・
LUT Library
CT&MR_KC_B
CT&MR_LOW_B
CT&MR_HIGH_B
CT&MR_KC_C
CT&MR_LOW_C
CT&MR_HIGH_C
LINEAR
DSA_KC_B
DSA_LOW_B
DSA_HIGH_B
DSA_LINEAR_B
DSA_KC_C
DSA_LOW_C
DSA_HIGH_C
DSA_LINEAR_C
US_KC_B
USER 4
USER 3
USER 2
USER 1
USER 4
USER 3
USER 2
USER 1
USER 4
USER 3
USER 2
USER 1
USER 4
USER 3
USER 2
USER 1
DR_HIGH_C
Library
x33
(unable to edit)
User LUT
x4
(able to edit)
(1)Copy the LUT of library
onto User LUT
(2) Edit the User LUT
curve point by point.
(3) Apply LUT in LUT
Library to each CH
of DRYPRO 771
(5) Setting the DICOM PRIORITY
Uses the standard LUT when the LUT is not
specified by the external diagnostic device.
LUT
CH1
CH2
CH3
CH4
Image Processing
Specifies the LUT No. registered on
DRYPRO 771 for image processing
External Diagnostic Device
1
2
3
4
5
6
7
1
2
3
4
5
6
7
1
2
3
4
5
6
7
1
2
3
4
5
6
7
U5I Press enter .
Copy from library
U52 LUT1 020-300 .
POINT01 020 X0.01D
U60 LUT LIBRARY /.
CH 1 MRI-1
S31 DEFAULT LUT .
DICOM
(4) Select the LUT for each CH.
U31 LUT NO. .
1 USER1
Select which to use
Default for seven LUTs each for
CH1~4 are all linear.
P-CON
DICOM
CH1
CH2
CH3
CH4
Use web maintenance tool to copy the
User LUT from one CH to other.
(1)
Copying the Library LUT to User LUT
When it is desired to manipulate and use the library LUT that is prepared by the
manufacturer, copy the library LUT to the User LUT.
(same operation can be implemented using "web maintenance tool
→
SCU
→
USER LUT")
(2)
Editing the User LUT curve point by point.
Editing can be done either by changing the points of the LUT copied from the LUT Library
or create a new LUT by inputting each point for the curve.
(same operation can be implemented using "web maintenance tool
→
SCU
→
USER LUT".
Using the web maintenance tool allows the operator to manipulate the LUT while viewing it
on the monitor)
(3)
Applying the LUT in the LUT Library to each CH of the DRYPRO 771.
Apply the LUT from the LUT Library which is provided by the manufacturer or the User LUT
to the channel 1~7 through which the corresponding external device is affected.
(same operation can be implemented using "web maintenance tool
→
SCU
→
LUT")
(4)
Selecting the standard LUT which will be used by each channel.
Set the standard LUT that will be used when the LUT is not specified by the external device
or when "S31DEFAULT LUT" is set "P-CON".
(same operation can be implemented using "web maintenance tool
→
SCU
→
DEFAULT
SETUP)
(5)
Setting the DICOM PRIORITY
Select between the LUTs used for image processing, one for standard LUT (P-CON) set for
DRYPRO 771 and the other specified for each image processing by the external device
(same operation can be implemented using "web maintenance tool
→
SCU
→
PRIORITY
SETUP)
p.69
p.70
p.71
p.73
p.75
Outline of LUT Setting
Outline of LUT generated on the DRYPRO 771 is described below.
3.7
Setting the Print Quality (U50, U60, U30)
How to set the quality of the print image of DRYPRO 771 is set out below.
Summary of Contents for DRYPRO 771
Page 10: ...Preface 8 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Page 26: ...Ch 1 Before Installation 24 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Page 54: ...Ch 2 Unpacking and Installation 52 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Page 82: ...Ch 4 Checking the Image 80 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Page 96: ...Ch 4 Checking the Image 94 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...