Appendix E: IEEE-488 conformance
Model 6517B Electrometer Reference Manual
E-4
6517B-901-01 Rev. C / August 2015
Coupled commands
Command Also
changes
To
If
:TRIG:COUN
:TRAC:POIN:AUTO
:TRAC:POIN
:TRAC:FEED:PRET:AMO:READ
OFF
TRIG:COUN
TRIG:COUN *
TRAC:FEED:PRE
T:AMO:PERC /
100
TRAC:POIN:AUTO
is
ON
and
TRIG:COUN
is
an illegal buffer size
TRAC:POIN:AUTO
is
ON
TRAC:POIN:AUTO
is
ON
:TRAC:FEED:CONT :TRIG:SOUR
TIM
TRAC:FEED:CONT
is
PRET
and
TRIG:SOUR
is neither
IMM
nor
TIM
:TRAC:POIN :TRAC:POIN:AUTO
:TRAC:FEED:CONT
:TRAC:FEED:PRET:AMO:READ
OFF
NEV
TRAC:POIN * TRAC
:FEED:PRET:AM
O :PERC / 100
:TRAC:FEED:PRET:
AMO:PERC
:TRAC:FEED:PRET:AMO:READ
:TRAC:FEED:CONT
TRAC:POIN * TRAC
:FEED:PRET:AM
O :PERC / 100
NEV
:TRAC:FEED:PRET:
AMO:READ
:TRAC:FEED:PRET:AMO:PERC
:TRAC:FEED:CONT
TRAC:FEED:PRET:
AMO:READ /
TRAC :POIN *
100
NEV
:TRAC:CLE :TRAC:FEED:CONT
NEV
...:APER
...:APER:AUTO
...:NPLC:AUTO
...:NPLC
...:DIG
OFF
OFF
APER*
line-frequency
varies per function
...DIG:AUTO
is
ON
...:NPLC
...:APER:AUTO
...:NPLC:AUTO
...:APER
...:DIG
OFF
OFF
NPLC/line-frequency
varies per function
...DIG:AUTO
is
ON
...:DIG
...:DIG:AUTO
...:APER
...:NPLC
OFF
varies per function
varies per function
...APER:AUTO
is
ON
...NPLC:AUTO
is
ON
...:RANG:UPP
...:RANG:AUTO
OFF
...:REF:ACQ
...:REF
presently displayed
reading
...:DET:FUNC
...:APER
...:DIG
...:AVER:STAT
...:AVER:COUN n
...:AVER:TCON n
...:AVER:ADV:STAT n
...:AVER:ADV:NTOL n
varies per function
varies per function
varies per function
varies per function
varies per function
varies per function
varies per function
...APER:AUTO
is
ON
...DIG:AUTO
is
ON
...AVER:AUTO
is
ON
...AVER:AUTO
is
ON
...AVER:AUTO
is
ON
...AVER:AUTO
is
ON
...AVER:AUTO
is
ON