![background image](http://html2.mh-extra.com/html/keithley/6517b/6517b_reference-manual_4068717419.webp)
Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-168
6517B-901-01 Rev. C / August 2015
:LLEVel <NRf>
This command specifies the low-level voltage for the square wave sweep test.
Usage
:TSEQuence:SQSWeep:LLEVel <NRf>
:LLEVel?
<NRf>
-100.0
to
100.0
Volts; 100V range
-1000
to
1000
Volts; 1000V range
Details
This configuration command is used for the square wave sweep test.
:LLEVel?
queries low-level voltage.
Also see
None
:LTIMe <NRf>
This command specifies the low-level time for the square wave sweep test.
Usage
:TSEQuence:SQSWeep:LTIMe <NRf>
:LTIMe?
<NRf>
0 s
to
9999.9 s
Details
This configuration command is used for the square wave sweep test.
:LTIMe?
queries low-level time.
Also see
None