Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-77
:COUNt <n>
This command is used to specify the filter count for volts, amps, ohms, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:AVERage:COUNt <n>
[:SENSe[1]]:CURRent[:DC]:AVERage:COUNt <n>
[:SENSe[1]:RESistance:AVERage:COUNt <n>
[:SENSe[1]]:CHARge:AVERage:COUNt <n>
:COUNt?
:COUNt? DEFault
:COUNt? MINimum
:COUNt? MAXimum
<n>
1 to 100:
specify filter count
DEFault:
10
MINimum:
1
MAXimum:
100
Details
These commands are used to specify the filter count. In general, the filter count is the number of
readings that are acquired and stored in the filter buffer for the averaging calculation. The larger the
filter count, the more filtering that is performed.
Also see
None
[:STATe] <b>
This command is used to control the filter for volts, amps, ohms, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:AVERage[:STATe] <b>
[:SENSe[1]]:CURRent[:DC]:AVERage[:STATe] <b>
[:SENSe[1]:RESistance:AVERage[:STATe] <b>
[:SENSe[1]]:CHARge:AVERage[:STATe] <b>
[:STATe]?
<b>
0 or OFF:
Disable the digital filter
1
or ON:
Enable the digital filter
Details
These commands are used to enable or disable the digital averaging filter for the specified function.
When enabled, readings are filtered according to how the filter is configured.
Attempting to set
STATe
to
ON
with
TYPE
at
NONE
and
MED:STAT
at
OFF
generates a "Settings
Conflict Error." Setting
STATe
to
ON
during a scan or test sequence changes
TCON
to
REPeat
.
Also see
None