Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-28
6517B-901-01 Rev. C / August 2015
Command Description
Default
parameter
SCPI
:ILINe <NRf>
Select input line (1 to 6)
2
:ILINe?
Query input line
:OLINe <NRf>
Select output line (1 to 6)
1
:OLINe?
Query output line
:SSYNchronous
Path to configure semi-synchronous
trigger link:
:LINE <NRf>
Select trigger line (1 to 6)
1
:LINE?
Query trigger line
Notes:
1. Defaults for continuous initiation:
:SYSTem:PRESet
enables continuous initiation.
*RST
disables continuous initiation.
2.
TLINK
and
RTCLock
(Arm Layer 1) are non-SCPI parameters for the
:SOURce
command. The
command (
:SOURce
), and all the other parameters are SCPI confirmed.
3. Defaults for count (Arm Layer 2 and trigger layer):
:SYSTem:PRESet
sets the count to INF (infinite).
*RST
sets the count to 1.