:PRESet
This command is used to return registers to their default conditions.
Usage
:STATus:PRESet
Details
When this command is sent, the SCPI event registers are affected as follows:
1. All bits of the positive transition filter registers are set to one (1)
2. All bits of the negative transition filter registers are cleared to zero (0)
3. All bits of the following registers are cleared to zero (0):
a. Operation event enable register
b. Questionable event enable register
c. Measurement event enable register
4. All bits of the following registers are set to one (1):
a. Trigger event enable register
b. Arm event enable register
c. Sequence event enable register
Registers not included in the above list are not affected by this command.
Also see
None