Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-144
6517B-901-01 Rev. C / August 2015
:ELEMents <item list>
This command is used to select elements for the buffer.
Usage
TRACe:ELEMents <item list>
:ELEMents?
<item list>
TSTamp:
includes time stamp
HUMidity:
includes humidity
CHANnel:
Includes channel number
ETEMperature:
Includes external temperature
VSOurce:
Includes V-source
NONE:
None of the above elements
Items in the list must be separated by commas.
Details
READing, STATus, RNUMber
(reading number) and
UNIT
are always enabled for the buffer and
are included in the response for the query (
:ELEMents?
).
This command is used to include additional elements for each buffer reading. The
:ELEMents
<item list>
command (under
FORMat subsystem
(on page 14-50)) provides a detailed
description for each element.
Also see
:ELEMents <item list>
(on page 14-53)
:TRIGger subsystem
The trigger subsystem is made up of a series of commands and subsystems to configure the three
layers of the trigger model (see Trigger models for more information) These commands and
subsystems are summarized in
:TRIGger command summary
(on page 14-26).