Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-159
:TLIne <NRf>
These commands specify and query the trigger link line.
Usage
:TSEQuence:TLIne <NRf>
:TLINe?
<NRf>
1
: Line #1
2
: Line #2
3
: Line #3
4
: Line #4
5
: Line #5
6
: Line #6
Details
This command is used to specify the input trigger link line for the TLINk control source. This
command is not used if a different control source is selected.
:TLINe?
queries the trigger link line.
Also see
:TSOurce <name>
(on page 14-158)
:STARt <NRf>
This command specifies the start voltage for a diode leakage or staircase sweep test.
Usage
:TSEQuence:DLEakage:STARt <NRf>
:TSEQuence:STSWeep:STARt <NRf>
:STARt?
<NRf>
-100.0
to
100.0
Volts; 100V range
-1000
to
1000
Volts; 1000V range
Details
This configuration command is used for the following tests:
Diode leakage test
Staircase sweep test
:TSEQuence:DLEakage:STARt
specifies the start voltage for a diode leakage test.
:TSEQuence:STSWeep:STARt
specifies the start voltage for a staircase sweep test.
:STARt?
queries the start voltage
Also see
None