Model 6517B Electrometer Reference Manual
Section 12: Common commands
6517B-901-01 Rev. C / August 2015
12-3
Common command descriptions
*CLS (clear status)
This command clears status registers and the error queue.
Details
The
*CLS
command is used to clear (reset to 0) the bits of the following registers in the Model 6517B:
Standard event status register
Operation event register
Error queue trigger event register
Sequence event register
Arm event status register
Measurement event register
Questionable event register
This command also forces the instrument into the operation complete command idle state and
operation complete query idle state.
Also see
None