Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-122
6517B-901-01 Rev. C / August 2015
DISable <list>
This command is used to disable messages for the error queue.
Usage
:STATus:QUEue:DISable <list>
<list>
(numlist)
Details
On power-up, all error messages are enabled and go into the error queue as they occur. Status
messages are not enabled and do not go into the queue. This command is used to specify which
messages you want disabled. Disabled messages are prevented from going into the error queue.
Messages are specified by numbers (see
Status and error messages
(on page 2-11)).
Also see
ENABle <list>
(on page 14-121)
:CLEar
This command is used to clear the error queue.
Usage
:STATus:QUEue:CLEar
Details
This action command is used to clear the error queue of messages.
Also see
None
:SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in
:SYSTem
command summary
(on page 14-23).
:PRESet
This command returns the instrument to states optimized for front-panel operation.
Usage
:SYSTem:PRESet
Details
This command returns the instrument to states optimized for front panel operation. :SYSTem:PRESet
defaults are listed in the SCPI tables (refer to the
SCPI command summary
(on page 14-6)).
Also see
None