Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-98
6517B-901-01 Rev. C / August 2015
:STIMe <n>
This command is used to specify the setting time.
Usage
:ROUTe:SCAN:STIMe <n>
:EXTernal?
:EXTernal? DEFault
:EXTernal? MINimum
:EXTernal? MAXimum
<n>
0.003 to 99999.999:
Specify settling time in seconds
DEFault:
0 seconds
MINimum:
0 seconds
MAXimum:
99999.999 seconds
Details
This command is used to specify the settling time for the relay switches on the internal scanner card.
After each switch closes, a measurement does not occur until after the specified settling time expires.
Also see
None
:SMEThod <name>
This command is used to select the switching method.
Usage
:ROUTe:SCAN:SMEThod <name>
:SMEThod?
<name>
VOLTage:
Break-before-make switching method
CURRent:
Normal switching method
Details
This command is used to select the switching method during an internal scan. The
VOLTage
method
ensures that each switch opens (breaks) before the next switch in the scan closes (makes). This
method ensures that two or more channels cannot be closed at the same time.
Use the
CURRent
switching method for optimum speed in applications where "break-before-make" is
not necessary.
Also see
None