Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-146
6517B-901-01 Rev. C / August 2015
:POFLag <name>
This command is used to control the no-operation-pending flag.
Usage
:INITiate:POFLag <name>
:POFLag?
<name>
INCLude:
Include no-operation-pending flag
EXCLude:
Exclude no-operation-pending flag
Details
This command is used to include or exclude the no-operation-pending flag in the internal process of
determining whether all pending operations of an initiate command are completed.
Also see
*OPC (operation complete)
(on page 12-8)
*OPC? (operation complete query)
(on page 12-9)
:ABORt
This command is used to abort the operation.
Usage
:ABORt
Details
When this action command is sent, the Model 6517B aborts the operation and returns to the top of
the trigger model. If continuous initiation is disabled, the instrument goes the idle state. If continuous
initiation is enabled, operation continues on into Arm Layer 1.
The abort command resets the source bypasses of the trigger model. If scanning, the scan pointer
resets back to the first channel in the scan list.
Also see
None