Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-63
[:UPPer] <n>
This command is used to set the measurement range for volts, amps, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:RANGe[:UPPer] <n>
[:SENSe[1]]:CURRent[:DC]:RANGe[:UPPer] <n>
[:SENSe[1]]:CHARge:RANGe[:UPPer] <n>
:RANGe[:UPPer]?
:RANGe[:UPPer]? DEFault
:RANGe[:UPPer]? MINimum
:RANGe[:UPPer]? MAXimum
<n>
0 to 210:
Expected reading in volts
0 to 21e-3:
Expected reading in amps
0 to 2.1e-6:
Expected reading in coulombs
DEFault
21e-3 (amps)
2.1e-6 (coulombs)
MINimum:
0 (all functions)
MAXimum:
Same as
DEFault
Details
This command is used to manually select the measurement range for the specified measurement
function. The range is selected by specifying the expected reading as an absolute value. The Model
6517B then goes to the most sensitive range that accommodate the expected reading. For example,
if you expect a reading of approximately 10 mA, simply let the parameter (
<n>
) = 0.01 (or 10e-3) in
order to select the 20 mA range.
Also see
None