Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-86
6517B-901-01 Rev. C / August 2015
:FSELect <name>
This command is used to specify the test fixture.
Usage
[:SENSe[1]]:RESistance:RESistivity:FSELect <name>
:FSELect?
<name>
M8009:
Model 8009 Resistivity Test Fixture
USER:
User-supplied test fixture
Details
This command is used to specify which test fixture is being used for resistivity measurements.
M8009
specifies the Model 8009 test fixture. When using the Model 8009, the only parameter that needs to
be entered is the sample thickness for volume resistivity.
Specify
USER
when using any other resistivity test fixture. When using this type of test fixture, you
have to be enter the other parameters for resistivity measurements.
Also see
:STHickness <NRf>
(on page 14-86)
:USER commands
(on page 14-87)
:STHickness <NRf>
This command is used to specify the sample thickness for volume resistivity.
Usage
[:SENSe[1]]:RESistance:RESistivity:STHickness <NRf>
:STHickness?
<NRf>
0.0001 to 99.9999:
Sample thickness in millimeters
Details
This command is used to specify the thickness (in millimeters) of the sample for volume resistivity
measurements. Sample thickness is not required for surface resistivity.
Also see
None