1-28
Return to
4200-900-01 Rev. K / February 2017
Section 1: Getting Started
Model 4200-SCS User’s Manual
SCP2 (Oscilloscope)
Keithley Instruments offers two scope cards: 4200-SCP2HR and 4200-SCP2. However, only one
scope card at a time can be used in the 4200-SCS system.
The scope card is a modular, dual-channel, high-speed digital storage oscilloscope (DSO). It uses
a high-speed memory digitizer (DC to 700 MHz) and an embedded digital signal processor (DSP).
The scope card has two input channels to capture and analyze a variety of time-varying signals.
KScope, a soft front-panel software used to view pulse waveforms, is included with the
4200-SCP2HR and 4200-SCP2 (see the
“How to Generate Basic Pulses”
in Section 5 for details).
KScope provides full control of the DSO and allows export of waveform data that is compatible
with Microsoft
®
Excel
®
. The plug-and-play drivers can be used in most application programming
environments (for example, National Instruments LabVIEW™, Microsoft® Visual Basic
®
, and
Microsoft
®
Visual C++
®
).
The primary differences between the two scope cards are ADC resolution and sample rate:
•
4200-SCP2HR
: 16-bit resolution, single-channel sampling rate of 400-million samples per
second (400 MS / s)
•
4200-SCP2
: 8-bit resolution, single-channel sampling rate of 2.5 billion samples per second
(2.5 GS / s)
Basic pulse characteristics of the two scope cards are listed in
. See the supplied ZTEC
User’s Manual for complete specifications of the scope card.
1
NOTE
Refer to the
Reference Manual, Pulse source-measure connections, page 11-38
details about scope card connectors and connections to DUT.
1. The ZTEC User’s Manual is on your Model 4200-SCS Complete Reference Product Information CD.
Table 1-2
Scope card characteristics
Scope card
characteristic
Specifications
4200-SCP2HR (ZT410-50K)
4200-SCP2(ZT450-50K)
Dual channel
Simultaneous sampling of both channels
Simultaneous sampling of both channels
ADC resolution
16-bit
8-bit
Bandwidth
50
Ω
: DC to 250 MHz
1 M
Ω :
DC to 125 MHz
50
Ω
: DC to 1 GHz
1 M
Ω :
DC to 350 MHz
Maximum input
50
Ω
: ±5 V DC
1 M
Ω :
±25 V DC
(derated 20 dB / decade above 10 MHz)
50
Ω
: ±5 V DC
1 M
Ω :
±150 V DC
(derated 20 dB / decade above 1 MHz)
Coupling
DC or AC
DC or AC
AC coupling
50
Ω
: 200 kHz high-pass
1 M
Ω :
10 Hz high-pass
50
Ω
: 200 kHz high-pass
1 M
Ω :
10 Hz high-pass
Probe attenuation
0.9 to 1000:1
0.9 to 1000:1
Analog filter
N/A
20 MHz or bypass
Total memory
Up to 1 M S per channel
Up to 2 M S per channel (1 channel interleaved)
Up to 1 M S per channel
Up to 2 M S per channel (1 channel interleaved)
Sample (S) rate
10 k S / s to 200 M S / s
10 k S / s to 400 M S / s (1 channel interleaved)
2.5 k S / s to 1.25 G S / s
2.5 k S / s to 2.5 G S / s (1 channel interleaved)
Acquisition time range
250 ns to 3355 s
50 ns to 419 s (2 M sample memory)