4200-900-01 Rev. K / February 2017
Return to
3-69
Model 4200-SCS User’s Manual
Section 3: Common Device Characterization Tests
Overview of the cycling-related tabs
When you double-click the name of a subsite, KITE displays the tabs shown in
. Use the Subsite Setup tab to configure cyclical tests, the Subsite
Data tab to view test results numerically, and the Subsite Graph tab to view
results graphically.
Figure 3-50
General tab overview
Configuring subsite cycling
Figure 3-51
Subsite setup tab
Use to enable or disable cycling. If
Enable Cycles
is
unchecked, KITE ignores the
Subsite Setup
tab
settings and executes the Subsite Plan without
Use to choose one of the following cycling modes:
•
Stress
↔
measure cycles: for example, HCI tests: by selecting a stress / measure
mode.
•
Only measure cycles: no stressing: by selecting
Cycle Mode
.
When
Cycle Mode
is
selected, enter here
the fixed number of
times (cycles) that
you want the subsite
to execute (up to 128,
maximum). No
stressing is
performed when in
When a stress /
measure mode is
selected, use to
apply and save the
settings you have
entered in the Setup
tab.
This button also
applies and saves
settings made in the
Device Stress
Properties window
(see
).
When
Stress /
Measure Mode
is
selected, use to open
the Device Stress
Properties window,
which is used to
configure the subsite
stressing
When
Stress/Measure
Mode
is selected, use to
set up cycle timing:
•
Linear,
if you want all
stress times to be the
same
•
Log
, if you want the
stress time to increase
logarithmically with each
successive cycle.
•
Log
, if you want to
specify each time.
When
Stress / Measure
Mode
AND
Log
cycle times
are selected, use to “ask”
KITE to stop stressing and
make measurements at the
intervals specified in the
Rate (s)
field:
in addition to
any intervals specified in
the
Stress / Measure
Cycle Times
area.
For
more information, refer to
Step D: Set periodic test
intervals (stress /
measure mode, log timing
only) on page 3-72
.
When
Stress/Measure
Mode
is selected, use to
set up cycle timing:
•
Linear,
if you want all
stress times to be the
same
•
Log
, if you want the
stress time to increase
logarithmically with each
successive cycle.
•
List
, if you want to
specify each time.
Summary of Contents for 4200-SCS
Page 2: ......
Page 4: ......