Section 1: Getting Started
Model 4200-SCS User’s Manual
SMU terminals and connectors
The locations and configuration of the 4200-SMU and 4210-SMU terminals are shown in
. Basic information about these terminals is summarized below. Refer to the
Connections and Configuration, page 4-1
, for additional information regarding SMU signal
connections.
WARNING
Asserting the interlock will allow the SMU and preamp terminals to become
hazardous, exposing the user to possible electrical shock that could result in
personal injury or death. SMU and preamp terminals should be considered
hazardous even if the outputs are programmed to be low voltage. Precautions
must be taken to prevent a shock hazard by surrounding the test device and
any unprotected leads (wiring) with double insulation for 250 V, Category I.
CAUTION
The maximum allowed voltage between COMMON and chassis ground is
±32 V DC.
Figure 1-9
Models 4200-SMU and 4210-SMU connectors
FORCE terminal
The FORCE terminal is a miniature triaxial connector used to apply the SMU FORCE signal to the
DUT when a preamp is not being used.
• The center pin is FORCE
• The inner shield is GUARD
• The outer shield is circuit COMMON
INSTRUMENT
CONNECTIONS
SMU ONLY
SMU AND GNDU
SENSE LO
GUARD
SENSE LO
COMMON
COMMON
FORCE
SENSE
GUARD
4210
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
PreAmp
Control
Connector
SENSE LO
Force
and
Sense
Connectors