Configure Device Stress Properties
To configure the device stress properties for the Segment / Stress Measure Mode
(see Figure 3-71):
Figure 3-71
Segment Stress / Measure Mode: Device Stress Properties
1.
Active Site
selection
:
• When active, this field is used to select the wafer site number.
• If there is only one wafer site, this field will be inactive.
2.
Pulser selection
:
• There is a Device Settings Properties window for each pulse generator card in the system
(Model 4225-PMU, 4220-PGU, 4205-PG2, or 4200-PG2).
• Select the pulser to be configured
.
3.
SMU Pins
:
•
No switch matrix
:
–
With no switch matrix, the active SMU pin fields must be set to
0
(no connection) or -
1
(high impedance).
–
The
-1
setting puts the SMU in a high impedance mode, which is necessary if it
shares a pin with a VPU.
•
Switch matrix
:
–
With a switch matrix added to the system, the pin number settings determine signal
routing for the SMUs through the matrix to the device pins.
4.
PG2 Matrix connections
:
• With a switch matrix added to the system, fields for PG2 Channel pins are active.
–
There is a PG2 pin connection setting each channel.
• The pin number settings determine signal routing for the PG2 Channels through the
matrix to the device pins.
2
4
3
5
6
7
1
9