4200-900-01 Rev. K / February 2017
Return to
4-33
Model 4200-SCS User Manual
Section 4: How to Control Other Instruments with the Model 4200-SCS
Second connect test
This connect UTM connects the device to the PGU and the GNDU. In the project navigator,
double-click the second
connect
test to open it.
shows the parameters that connect
the device to the PGU. Not shown is line 1 (OpenAll) that opens the relays closed by the previous
connect test. Line
1
is shown in
.
For the parameters shown in
, the device connection pathways to the PGU and GNDU
. Remember that if your physical matrix connections are different, you will
have to change the connection parameters in the UTM to match them.
Figure 4-49
Second connect test - connects the device to the PGU
Figure 4-50
Signal paths to apply the pulse stress
pgu1-init test
In the project navigator, double-click
pgu1-init
to open the test. This one parameter test (see
) initializes the PGU. For example, it disables the output, resets errors, and sets
triggering. More information about the initialized state is provided in the description area of the
definition tab. For details about the UTMs for the pulse generator, refer to the
Pulse generator test example, Appendix F
.
Figure 4-51
PGU initialization
Connects PGU to pin 5 of test fixture
Connects GNDU to pin 6 of test fixture
Gate
Source
D
rain
Substrate
N-C
h
annel
MOS
F
E
T
1
2
3
4
5
6
A
B
C
D
E
F
S
M
U2
S
M
U3
GN
D
S
M
U
4
PGU
L
O
S
M
U
1
Initializes HP 8110
Summary of Contents for 4200-SCS
Page 2: ......
Page 4: ......
Page 334: ...Index 6 4200 900 01 Rev K February 2017 Model 4200 SCS User s Manual Index...