4.3.15 Measuring power
Note:
In order to measure the power, both the device and the measuring equipment must be
operated in Standby mode for at least 30 mins. beforehand.
– Connect the device to the correct supply voltage.
– Connect the dual-pedal footswitch 200138 30.
– Start the device.
– Set the NE to the value “any”.
– Select any procedure.
– Set the modes, the power settings and effects in accordance with the test report.
– Use the appropriate test cables for the HF socket modules according to the device version:
Example 205352 20-115: Measuring cable with 4 mm lab socket for
bipolar round
MF socket
monopolar socket
NE jack
– The basic test configuration is represented in the Figure 4.9 below.
Figure 4.9
12.06 / V1.0
4-11
AUTOCON
®
II 400
MODEL 205352 20-1
Fixed
load
Foot
switch
Scope
Var.
load
F1
F3 F4
F2
F5
Metron QAES
STORZ
autocon II 400
205352 20
F
connections monopolar
connections bipolar
connections multifunctional (MF)
Summary of Contents for 205352 20-1
Page 8: ......
Page 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x ...
Page 10: ......
Page 13: ...III Geräteabbildungen Photographs of the equipment Imágenes del equipo ...
Page 14: ...IV W R Q Geräteabbildungen Photographs of the equipment Imágenes del equipo I O Y U T E ...
Page 146: ......
Page 152: ......
Page 156: ......
Page 158: ......
Page 172: ......
Page 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1 ...
Page 192: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 203: ......