121
Table 204
Guidance and manufacturer’s declaration – electromagnetic immunity –
for equipment and systems that are not life-supporting
AUTOCON
®
ll 400 model 20 5352 20-11x is intended for use in the electromagnetic environment specified below. The customer or user of the
AUTOCON
®
ll 400 model 20 5352 20-11x should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the AUTOCON
®
ll 400 model
20 5352 20-11x, including cables, than the recommended
separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
3 V
rms
[
]
P
d
3
/
5
.
3
=
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
[
]
P
d
3
/
5
.
3
=
80 MHz to 800 MHz
[ ]
P
d
3
/
7
=
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked with the
following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects and
people.
a Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the unit is used exceeds the
applicable RF compliance level above, the unit should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the unit.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Anexo
Indicaciones sobre compatibilidad
electromagnética (CEM)
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Appendix
Electromagnetic Compatibility (EMC)
Information
Summary of Contents for 205352 20-1
Page 8: ......
Page 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x ...
Page 10: ......
Page 13: ...III Geräteabbildungen Photographs of the equipment Imágenes del equipo ...
Page 14: ...IV W R Q Geräteabbildungen Photographs of the equipment Imágenes del equipo I O Y U T E ...
Page 146: ......
Page 152: ......
Page 156: ......
Page 158: ......
Page 172: ......
Page 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1 ...
Page 192: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 203: ......