SENSe:TEMPerature:RESolution?
Queries the temperature resolution.
Returns
1, 0.1, 0.01 or 0.001
SENSe:ZERO:AUTO {mode}
Turns the auto-zero mode on or off. Default at power-up is auto-zero
mode off.
{mode} 1,
on
Auto-zero
on
0, off
Auto-zero off
Auto-zero is only available for single channel measurements. The
'on' mode is automatically reset to 'off if any of the following
commands are sent:
CONFigure:TEMPerature:TC {type},{rj mode},{ext rj standard}
CONFigure:TEMPerature:RTD {type},{standard},{con},{current
mode},{root 2}
CONFigure:CHANnel {channel}
MEASure:TEMPerature:TC? {type},{rj mode},{ext rj standard}
MEASure:TEMPerature:RTD? {type},{standard},{con},{current
mode},{root 2}
MEASure:CHANnel? {channel}
ROUTe:SCAN:LSELect {slx}
SENSe:ZERO:AUTO?
Queries the auto-zero mode status.
Returns 1 Auto-zero
on
0 Auto-zero
off
Communications Interface
Page 7-21
Summary of Contents for TTI-7-R