Communications Interface
On completion of the scanner routine the next DATAIogger:STEP
command will re-set the scanner routine to the beginning and start
the next cycle.
If the data logger memory is full, any further DATAIogger:STEP
commands will generate an "execution error" the command will be
ignored.
Terminates the current measurement cycle.
Sets the trigger mode to Single shot.
Initiates a single measurement and stores the result in the next
location of the data logger memory.
DATAIogger:VALue? {<value>}
Transfers a specified number of data logger reading from the
memory to the output buffer.
Sending ABORt will terminate the operation.
{value} 1 to 4000 Request a specific number of readings
all
Request all data logger readings
The output format is:
{<recordno.>},"{channel}",{<temperature>},"{unit}","{date}", "{time}"
DATAIogger:POINts?
Query the number of readings actually stored in the data log
memory.
Returns
0
No readings stored
4000 Maximum
readings stored
Page 7-33