9.8 Display
User interface, display
40x 8 character, (240x 64 pixel)
NTN LCD green/ yellow back lit display
9.9 Physical
Dimensions:
Metal case with adjustable rest/handle,
240mm deep (front to back) x 260mm wide x 80mm high
(9.45" deep x 10.24" wide x 3.15" high)
Weight:
5.5kg (12.1 Ibs)
9.10 Pt25 and 100 System accuracy
System accuracy is the combined accuracy of the TTI-7-R plus the calibration uncertainty of
the PRT. Accurate figures for the instrument and system with the range of calibrated PRTs
available for the TTI-7-R are shown below.
TTI-7-R + WSP660 (-190 °C to + 660°C)
-180°C -50 °C
0°C
+100°C
+250 °C +660
°
C
±30mK ±20mK
±20mK
±20mK
±20mK ±25mK
Specifications
9.7 Analogue output option
Analogue output port
-2V to +2V range, 1 mV/°C scaling, 1°C resolution