Operating the TTI-7-R
5.7.3.2 Configuring the IEEE 488.2 communications interface
If an IEEE-488.2 parallel communications interface card is fitted,
select the
[Rem I/F]
option in the main configuration menu.
The screen displays:
Talk
Only
Addr Quit OK
Select the
[Addr]
option, the screen displays the current selected
IEEE address.
Instrument address is 7
Chg
Quit OK
To change the address select [Chg] and enter the new address
from the function key pad.
Addresses in the range 0 to 30 are valid addresses.
The default factory set address is 7.
5.7.4 Setting up Trigger Mode
There are two trigger modes available to control measurement and
data logging operations:
[Run]
Run/Hold mode, continuous measurement option.
Pressing 3/TRIG in this mode alternates between
continuous measurement and measurement hold. The
power up default state is continuous measurement
operation.
[Sng]
Single shot mode. Single measurement mode collects a
single measurement each time you press the 3/TRIG key
and pauses between key presses.
Page 5-30
Summary of Contents for TTI-7-R
Page 16: ...Setting up the TTI 7 R This page has been left blank intentionally Page 2 4...
Page 34: ...Measuring Temperature This page has been left blank intentionally Page 4 8...
Page 66: ...This page has been left blank intentionally Page 5 32 Operating the TTI 7 R...
Page 68: ...Options and accessories This page has been left blank intentionally Page 6 2...
Page 116: ...Specifications This page has been left blank intentionally Page 9 4...
Page 118: ...Communications Interface This page has been left blank intentionally Page 10 2...
Page 120: ...This page has been left blank intentionally Page 11 2...